20
21
The parameters of the presets can also be
changed make it different from default setting.
If you want to change back to default setting,
long press Setting button “P” to enter preset
program during standby state, select a preset
and press "Main" button to confirm, the default
setting will be reloaded. The preset program can
also be restored to default settings by turning off
the motor handpiece and powering back on.
Changing the preset program default setting is
not recommended, otherwise there is risk of file
breakage.
4.5 Handpiece functions setting
With the motor handpiece turned off, hold down the Setting button “P”
and press Main button to enter handpiece function settings, press Setting
button “P”
to cycle through settings, press Adjusting button “+”/“-” to
adjust, then press Main button to confirm.
With the motor handpiece turned off, hold down
the Setting button “P” and press Main button to
enter
handpiece function settings, the software
version number will appear on the display
screen.
After 3 seconds of displaying the version
number on the screen, the "Auto Power OFF"
can be changed, press Adjusting button “+”/“-
” to adjust, then press the "Main" button to
confirm.
This is the auto power off time of motor
handpiece if no buttons are pressed. It can be set
from 3 to 30 minutes in 1 minute increments.
Press Setting button “P” again, the “Auto
Standby Scr” can be changed, press Adjusting
button “+”/“-” to adjust, then press to "Main"
button to confirm.
This is the automatic return to standby display
of motor handpiece when no buttons are
pressed. It can be set from 3 to 30 seconds in 1
second increments.
Press Setting button “P” again, the "Dominant
Hand" can be changed, press Adjusting button
“+”/“-” to adjust, then press to “Main” button to
confirm. The right hand and the left hand can be
set.
Press Setting button “P”again, the “Calibration”
can be changed, press Adjusting button “+”/“-”
to select “ON”, then press "Main" button for
calibration. Before calibrating, make sure the
original contra angle is installed, and do not
install the file. The torque will not be correct if
it is calibrated without original contra angle or
if there is any load on contra angle chuck, and
there is risk of file damage.
After replacement of contra angle, the contra
angle should be calibrated before use.
Press Setting button “P”again, the "Beeper
Volume" can be changed ,press Adjusting button
“+”/“-” to adjust, then press to "Main" button to
confirm.
The"Beeper Volume"can be set from 0-3.
Vol.0: Mute.
Press Setting button “P” again, the "Restore
Defaults" can be changed, press Adjusting
button “+”/“-” to select “ON”, then press to
"Main" button to restore defaults.
4.6 Protective function of automatic reverse
During operation, if the load value exceeds the preset torque value,
the file rotation mode will automatically change to Reverse Mode. And