Axetris AG
LGD F200P2 Operating Instructions - Rev D
19.09.2018
51/60
Example 3 [50 ppm = 8 mA or 1.25 V; 100 ppm = 16 mA or 3.75 V]
min
max
delta
Concentration [ppm]
50
100
50
Analog voltage [V]
1.25
3.75
2.5
Analog current [mA]
8
16
8
normalized []
0.25
0.75
0.5
01
.
0
50
5
.
0
ppm
Span
25
.
0
25
.
0
01
.
0
50
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_
Offset:
AnalogOut_
Span:
50
100
-0.25
0.01
Example 4 [50 ppm = 16 mA or 3.75 V; 100 ppm = 8 mA or 1.25 V]
min
max
delta
Concentration [ppm]
50
100
50
Analog voltage [V]
3.75
1.25
-2.5
Analog current [mA]
16
8
-8
normalized []
0.75
0.25
-0.5
01
.
0
50
5
.
0
ppm
Span
25
.
1
75
.
0
)
01
.
0
(
50
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_
Offset:
AnalogOut_
Span:
50
100
1.25
-0.01