Axetris AG
LGD F200P2 Operating Instructions - Rev D
19.09.2018
50/60
C. Examples for Analog Output Adjustments
See Ch. 2.12.5.2 for instructions.
Example 1 [0 ppm = 4 mA or 0 V; 100 ppm = 20 mA or 5 V]
min
max
delta
Concentration [ppm]
0
100
100
Analog voltage [V]
0
5
5
Analog current [mA]
4
20
16
Normalized (N
1
; N
2
) []
0
1
1
01
.
0
100
1
ppm
Span
0
0
01
.
0
0
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_
Offset:
AnalogOut_
Span:
0
100
0
0.01
Example 2 [50 ppm = 4 mA or 0 V; 100 ppm = 20 mA or 5 V]
min
max
delta
Concentration [ppm]
50
100
50
Analog voltage [V]
0
5
5
Analog current [mA]
4
20
16
normalized []
0
1
1
02
.
0
50
1
ppm
Span
1
0
02
.
0
50
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_
Offset:
AnalogOut_
Span:
50
100
-1
0.02
Содержание LGD F200P2
Страница 1: ...Operating Integration Instructions Laser Gas Detection OEM Module LGD F200P2 Axetris AG...
Страница 57: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 57 60 Notes...
Страница 58: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 58 60 Notes...
Страница 59: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 59 60 Notes...
Страница 60: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 60 60 Notes...