Axetris AG
LGD F200P2 Operating Instructions - Rev D
19.09.2018
35/60
3.5 Data transmission protocol
The LGD sensor has multiple commands to communicate with other devices, like microcontrollers, or with
the customer user interface when it is connected to a PC. The following describes the structure of all
commands necessary for the integration and use of the sensor.
RS232 Serial settings
Default parameters for the serial
communication to the sensor:
Baud rate
9600 (19200, 115200)
Data bits
8
Parity
no
Stop bit
1
Flow control
no
Overall communication structure
In the following description we will use
Command,
describing the transfer from PC to the sensor.
Conversely, the name
Response
will denominate the equivalent transfer from the sensor to the PC.
The
Command
or
Response
IDs consist of the following:
Command / Response ID
Abbreviation
Ping
P
Config
C
Start Measurement
M
Idle
I
Save to EEPROM
S
Version.
V
Status monitoring
D
The data transfer is based on packets constituted of a multiple of 8 Bytes.