3.3 Thermal Management of the LGD Analyzer Module @220°C
When a device is used at 220°C / 428°F (cell temperature) extra care should be taken with the
thermal management.
To avoid having too high temperature on the laser assembly (that would lead to an error
and trigger the safe mode), the LGD F200P2P2-H NH3 220C device should not be mounted
vertically with the laser on top. The below figure explains which orientations are
acceptable and which orientations must be avoided.
All horizontal mounting positions are allowed. The electronic board can be on any side (top, side, bottom).
Vertical mounting is possible only if the cell heater screw terminal connector is not on top position