G01
SPD Owner’s Manual
30
Feed Conductivity - Product Conductivity
Rejection (%) =
Feed Conductivity
Since the system conductivity monitor displays the two in a 10 to 1 ratio, the approximate rejection is
available at a glance. If the needle on the meter does not move when the display is shifted from feed
to product it indicates the rejection is 90%. If the needle when indicating product is only half the
value of the feed, then rejection is 95%. If the needle when indicating product is only one fourth of
the value of the feed, then rejection is 97.5%. This is illustrated below for several different feed
waters.
Содержание SPD2H
Страница 2: ......
Страница 11: ...G01 SPD Owner s Manual 9 FIGURE 5 ...
Страница 22: ...G01 SPD Owner s Manual 20 FIGURE 7 ...
Страница 33: ...G01 SPD Owner s Manual 31 ...
Страница 35: ...G01 SPD Owner s Manual 33 Appendix 1 Keyence Ladder Program ...
Страница 36: ...G01 SPD Owner s Manual 34 ...
Страница 37: ...G01 SPD Owner s Manual 35 ...
Страница 38: ...G01 SPD Owner s Manual 36 ...