G01
SPD Owner's Manual
13
Some amount of disinfectant “rebound” will occur. This is when a disinfectant residual reappears after a
period of RO shutdown even though it was not present earlier. Rebound is minimized in the SPD RO
due to use of Circserts and the annulus drains, but still exists. This is because of disinfectant trapped in
the dead ended tubing to the pressure gauges, or anywhere that was not or cannot be properly rinsed.
Hazards of Leaks:
The recommended Minncare
TM
disinfectant is very corrosive to most metals and
several plastics. If any of the diluted disinfectant leaks to the outside at tubing connections, the spilled
disinfectant has the potential of destroying the fitting over time. So any leaks must be hand rinsed
copiously and dried.