In the above test setup, all cables, attenuators, and oscilloscope should have
bandwidth ratings of at least 2 GHz (and preferably > 10 GHz). The use of 40 dB
attenuator at the sampling scope vertical input channel will insure a peak input signal to
the sampling scope of less than 1 Volt, to avoid damaging the scope.
BASIC TEST ARRANGEMENT, WITH OUTPUT MODULES
For normal operation, one or both of the output modules will be installed on the outputs.
When using OUT1, the basic test arrangement is shown below:
21
TRIG
CONNECTOR
AC
POWER
ALL CABLES: 50 OHM COAXIAL
50 OHM INPUT
TRIG
CONNECTOR
40 dB
ATTENUATOR
AVTECH
AVR-3-PW-TEK3-B
OUT2
OUT1
SAMPLING
OSCILLOSCOPE
BW > 2 GHz
AVTECH
AVR-3-PW-TEK3-B
OUT1
SYNC
OUTPUT
OUT2
DIRECTION
OF CURRENT
IN
B
A
PROBE*
12" / 30 cm
RG-58
cable
OSCILLOSCOPE
CHANNEL A
TRIG
INPUT
6" LENGTH OF
RG-316 CABLE
(SUPPLIED)
AVTECH
AVX-TEK3-TM1
* Depending on the exact current probe model, a probe amplifier
may be needed between the probe and the oscilloscope.