The OUT1 and OUT2 connectors are different. OUT1 has a BNC connector, and OUT2
has an SMA connector. The output modules have matching connectors, to make it
obvious which output module is to be used with which output connector.
OUTPUT MODULE EQUIVALENT CIRCUIT
The output modules serve two functions: they provide a physically convenient location
to which to clamp the current probe under test to, and they contain the resistive load (or
termination) which is required to define the current amplitude. The equivalent circuit
(valid for both output modules) is shown below:
The connectivity of the ground shield of the shorting cable is deliberately broken so that
current can not flow in the shield - if the cable shield was active, it would prevent the
current probe from sensing the current on the center conductor!
23
IN
B
A
BREAK IN
GROUND SHIELD
CONTINUITY
(BETWEEN
CONNECTOR
AND CHASSIS)
VOLTAGE
PULSE
FROM
MAINFRAME
COAXIAL
CABLE
SHIELDS
(GROUNDED)
COAXIAL
CABLE
CENTER
CONDUCTOR
OUTPUT
MODULE
ALUMINUM
CHASSIS
(GROUNDED)
50Ω
LOAD
TO
GROUND
SHORTING
CABLE
SMA
CONNECTOR