MEASUREMENT ACCURACY AND CALIBRATION
The AVR-3-PW-TEK3-B is a voltage pulser. The output current is defined by the
voltage amplitude and the load resistance.
The voltage amplitude accuracy is specified to be the greater of ±2% or ±0.5V, when
shipped from the factory.
The load resistance is specified to be 50Ω ±1%, when shipped from the factory.
The as-shipped current amplitude accuracy can thus be estimated as ≈ √ (0.02
2
+
0.01
2
) = 2.2%.
VOLTAGE AMPLITUDE CALIBRATION
For applications requiring greater accuracy, the user can observe and calibrate the
voltage output. The commands required to update the internal voltage amplitude
calibration are described at http://www.avtechpulse.com/appnote/techbrief5, in the
Technical Brief entitled "How Can I Calibrate the Amplitude and Offset of my Avtech
Pulse Generator?".
LOAD RESISTANCE CALIBRATION
The user can also measure the load resistance between the center conductor of the "B"
connector and the outer shield of the "A" connector on each output module, with all
cables removed. Beware that the outer shield of the "B" connector is
not
connected to
ground. Four-wire measurements must be used for accurate results.