a) Press the MOVE button until the arrow pointer is pointing at the "AMP1" menu
item.
b) Press the CHANGE button. The amplitude submenu will appear. Rotate the
ADJUST knob until the amplitude is set at 250V.
c) Observe the oscilloscope. You should see 500 ns wide, 250V pulses.
d) Rotate the ADJUST knob. The amplitude as seen on the oscilloscope should
vary.
e) Press CHANGE to return to the main menu.
10. To change the Channel 2 output amplitude:
a) Press the MOVE button until the arrow pointer is pointing at the "AMP2" menu
item.
b) Press the CHANGE button. The amplitude submenu will appear. Rotate the
ADJUST knob until the amplitude is set at 50V.
c) Observe the oscilloscope. You should see 200 ns wide, 50V pulses. Due to the
very fast rise time of this output, some ringing will be observed. This is due to
limitations of the oscilloscope, not the pulser. A high-bandwidth sampling
oscilloscope may be used to confirm this, if desired.
d) Rotate the ADJUST knob. The amplitude as seen on the oscilloscope should
vary.
e) Press CHANGE to return to the main menu.
11. Try varying the pulse width, by repeating step (6). As you rotate the ADJUST knob,
the pulse width on the oscilloscope will change. It should agree with the displayed
value.
This completes the operational check.
35
Содержание AVR-3-PW-TEK3-B
Страница 31: ...Larger values of attenuation 6 dB will reduce the reflections and the pulse amplitude further 31 ...
Страница 42: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY ...
Страница 43: ...PCB 235B HIGH VOLTAGE DC POWER SUPPLY ...
Страница 44: ...PCB 284B HIGH VOLTAGE DC POWER SUPPLY WITH DISCHARGE ...
Страница 46: ...PCB 241B DUAL CHANNEL PULSE WIDTH BOARD ...
Страница 51: ...PERFORMANCE CHECK SHEET 51 ...