Pico Water Purification System Operation Manual
Installation and Setup
7
Tubing
These are general guidelines for connecting the tubing to the Pico connectors.
•
Use a sharp blade to cut the tubing
•
Be sure the tube is long enough to allow for disconnecting and reconnecting during service.
•
Be sure tubing is clean and not cracked
•
Be sure there are no obstructions in the tube before connecting
•
Do not use excessive force when inserting tubing. Excessive force may damage the tubing resulting in
leaks.
•
Pull on the tube only once to be sure it is secured in the fitting
•
After pulling once, insert the wrench under the collet and push the tube into the fitting once more for
a secure fit.
Connect High or Low Pressure Feed Water Tubing
NOTE:
Use the provided elbow connectors as necessary.
The
5/16-inch feed water tubing can be connected to the high pressure or low pressure inlet fitting
depending on the known feed water pressure.
NOTE:
If the pressure is not known, connect the feed water tubing to the high pressure fitting. A water pressure
test will be conducted at the completion of the installation process to determine the required pressure.
Follow this procedure.
1.
Measure and cut the length of the
5/16-inch
feed water tubing needed between the water source
connection and the applicable feed water inlet fitting on the Pico unit. Add extra length to accom-
modate removal and reconnection of the tube.
2.
Connect the check valve-end of the tubing to the applicable inlet fitting.
Feed water inlet. Low pressure, less than 1 bar [15 psi]
Feed water inlet, High pressure, greater than 1 bar [15 psi]
Содержание Pico PICO10T2
Страница 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Страница 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Страница 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Страница 38: ......
Страница 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...