Installation and Setup
11
9.
When the pressure displayed is less than 0.1 bar, remove the feed water tubing from the high pres-
sure inlet fitting and connect to the low pressure inlet fitting
.
10.
Turn on water to the Pico unit.
11.
After 3 minutes check the pressure displayed on the screen. The pressure must be greater than
0.6 bar.
12.
Press
OK
.
13.
After 2 minutes, using a flat-blade screwdriver, adjust the pump pressure screw
on the right
side of the unit until
6.0 bar
(±0.2 bar).
14.
Press
OK
.
Rinse RO Module
The RO portion of the combination Pre-Treatment and RO module must be rinsed before first use. If a
DI cartridge is required for Type 2 water, the bypass tube must be connected between the green RO water
outlet tube and the white reservoir inlet tube
IMPORTANT:
Do not connect the DI cartridge (if applicable) before the RO module is rinsed. Bypass tubing
must be connected between the green RO water outlet tube and the white reservoir inlet tube.
Follow this procedure.
1.
Take one of these actions.
2.
On the control display, press
.
3.
Touch
4.
Select
TC021/TC022
.
5.
When
Remove old TC021/TC022
appears, touch
6.
When instructed to insert new TC021/TC022, touch
. The module will rinse for one hour.
7.
When operation is complete, touch
.
8.
If the bypass tube was connected, disconnect the bypass tube. Store the bypass tube in a safe loca-
tion.
If a DI cartridge is...
Then...
Required
Be sure the bypass tube is connected. See Figure 6.
Not required
Содержание Pico PICO10T2
Страница 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Страница 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Страница 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Страница 38: ......
Страница 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...