Avidity Science, LLC
Pico Water Purification System Operation Manual
Maintenance
25
7.
Remove the DI cartridge from its packaging.
8.
Be sure the green and white tubes are connected according to the color labels on the cartridge.
Figure 16. DI cartridge tube connections.
9.
Secure the tubes into the clamps near the connections.
10.
Touch
11.
On the control display, press
.
12.
Press .
13.
Select
TC023
.
14.
Press
until
Rinse
is displayed.
15.
Press
OK
. The unit will run for 10 minutes and automatically stop when complete.
16.
When operation is complete, press firmly at the center of the timer label sticker until a thin red line
appears. See Figure 17.
Figure 17. Activate DI cartridge timer.
17.
Replace and secure the left front cover of the Pico unit.