Maintenance
27
18.
Check for leaks from the UV lamp connection at the reservoir.
19.
Take one of these actions.
20.
Drain the reservoir until the water level is below the UV lamp.
21.
Disconnect the power supply.
22.
Remove the right front panel.
23.
Disconnect the UV lamp from its power connector.
24.
While holding the socket on the outside of the tank, insert your hand into the reservoir and loosen
and remove the nut that secures the UV lamp.
25.
Again, while holding the lamp socket in place, secure the nut to the socket from inside the reser-
voir.
26.
27.
Secure the cap to the top of the reservoir.
28.
Secure the right cover with the four hex head screws and replace the left cover.
29.
Secure the left cover into position.
If a leak is...
Then...
Not present,
Present,
go to step 20.