16
ZQ375 Checkweigher User Instructions
Introduction
Weigh / Checkweigh mode
- Press
UNDER
to briefly display the active under
value. Press and hold
UNDER
to add or modify an existing under value tolerance.
Menu navigation
- N/A
Numeric / Tolerance Entry
- N/A
Weigh / Checkweigh mode
-
TARGET
key acquires a target value, when
applicable. Its function changes in different applications. See the appropriate
application section.
Menu navigation
- N/A
Numeric / Tolerance Entry
- N/A
Weigh / Checkweigh mode
- Press
OVER
to briefly display the active over value.
Press and hold
OVER
to add or modify an existing over value tolerance.
Menu navigation
- N/A
Numeric / Tolerance Entry
- N/A
Weigh / Checkweigh mode
- Press the
ID
key briefly to view the active ID number.
Press and hold the
ID
key to view a prompt for ID number entry. Use the Numeric
to scroll in a new ID.
Menu navigation
- N/A
Numeric / Tolerance Entry
- N/A
Weigh / Checkweigh mode
- Press to access PLU database, if enabled.
Press and hold to access the menu password display.
Menu navigation
- Press to escape a screen without doing anything and move up
in the menu.
Numeric / Tolerance Entry
- Press to escape a screen without doing anything and
move up in the menu.
UNDER
TARGET
OVER
ID
F1
EU trade regulations state a scale is not allowed to display more than 20 divisions
below gross zero. Any weight below this allowable limit will cause the display to show
under bars (_ _ _ _).
When a tare is active, EU trade regulations do not allow the scale to be zeroed. If the
scale displays under bars with no weight on the scale and the
ZERO
key will not re-
zero the scale, this could mean the zero was changed prior to setting a tare.
To clear this, press and hold the
TARE
key until CLEAr is displayed and then press
the
ZERO
key. The scale should now display gross zero.