70
ZQ375 Checkweigher User Instructions
Supervisor menu
3.
Press
or
UNITS
to toggle between the choices and when your choice is
displayed, press
ZERO
to accept …
EnAbLE
is displayed.
Timeout
Battery
Enable
Timeout
4.
Press
UNITS
to go to the next menu item …
tMout
is displayed. Use this to set the length of time before inactivity on
the scale and keypad cause battery power to be shutoff. Values between
1 and 3600 minutes are valid. Default value is 60 minutes.
5.
Press
SELECT
…
A numeric entry screen appears.
6.
Use the
Numeric entry procedure on page 19
to key an a value, in minutes and
press
ZERO
to accept …
tMout
is displayed.
7.
This completes the Supervisor menu for the Sim375 application. Repeatedly
press
TARE
to return to normal operating mode.
Only enable the battery and set the
tMout
value if the ZQ-BAT option is being used. If
battery use is enabled, setpoint output 3 cannot be used for setpoints in
checkweighing applications. It is used as a shutoff signal to the ZQ-BAT battery
option.
Non ZQ-BAT supplied power may require external shut off circuitry. See the Service
Manual for information on this feature.