C Code Error
Action
Message
C3
Platter Ref
mark
not
detected.
Check the wind count and Attempt to
recalibrate 3 times with no error. If
unsuccessful throw error and lock touch
screen.
“Ref Tab not detected on
platter X see supervisor”
Notes: When a C3 error occurs, the affected platter did not detect the magnetic indicator on the platter (MP) or
the chain. This can be caused by damage to the indicator or the magnetic pickup. If this happens the controller
can no longer store its wind count to volatile memory. The C3 error stops this from happening. If this occurs
one should check the ref mark indicator lights on the controller (red LED). The MP led should light when the
Ref tab passes the magnetic pickup.
On all other platters, the indicator light should stay on until the Ref tab passes magnetic or optical pickup. If
this does not occur, the machine may have wiring problem or a bad controller.
C Code Error
Action
Message
C4
Platter Ref
mark
was
detected but
not with the
Ref
error
limitation.
Check wind count and Attempt to
recalibrate 3 times with no error. If
unsuccessful throw error and lock touch
screen.
“Ref Tab not detected on
platter X see supervisor”
Notes: A C4 error is caused by a condition where the machine has rotated more than 84 pulses past where it
expected to see the Ref mark on the affected platter. This can be caused by a missing Ref tab or more likely the
use of the manual platter movement switches. If a user moves the platter in a direction that would move it pass
the ref tab and then the machine fetches a bin in the opposite direction that condition could occur where the
machine would detect more than the limiting pulses. The machine will normally self correct this condition by
recalibrating.
C Code Error
Action
Message
C5
Baffle
pot
voltage out
of range.
None then lock the touch screen. The
administrator must use manual switch to
move to position 1 check the rack and
pinion and then perform a “CPR B”
Command from the maintain screen.
“Baffle over traveled See
supervisor”
Notes: A C5 error normally occurs when the baffle motor overdrives the pinion on the rack. Once this
happens, the user must check the rack and pinion to see if any damage has occurred. If the pinion is still
engaged on the rack, the system administrator should use the manual baffle switch to position 1 or the 12
th
pie
configuration setting. Next, the administrator should use the Maintain screen on the touch screen to reset the
baffle pot voltage. The command used to do this is “CPR B” once this has been completed, the machine should
be restarted to force a calibration.