Water Cooler Instruction Manual 

Model: 826A 

© Aurora Scientific, 2015                                        VER. 1.00 

page 7 

 

 

Figure 2: Filling Coolant Reservoir 

NOTE: Be careful not to overfill the reservoir, since any overflow from the reservoir will drain 

directly into the 826A case.  You should fill the reservoir to approximately 1.5cm (1/2”) from the 

top of the reservoir.  If you should overfill the reservoir, screw the cap back on, place a towel under 
the liquid circulator and tilt the unit from left to right to drain the liquid from the case. Under no 
circumstances should you turn the 826A upside down to drain overflow liquid from the case since 
this will damage the fan due to the liquid draining through the fan.  . 

Once the reservoir has been filled, switch on the 825A to allow the pump to fill the rest of the system and 
tubing with coolant.  When the system begins running for the first time the fluid inside the reservoir will 
begin to drain.  As this happens continue to fill the reservoir with distilled water until the draining has 
stopped.  Once full, replace the cap and unplug the power adapter from the system until ready for use. 

If the fluid refuses to flow through the system immediately when power is connected, it is likely that the 
pump is still dry.  In this case you may attempt to prime the pump by slightly tilting the unit back and 
forth to encourage the liquid to flow. Once the pump has begun to circulate, you may replace the 
circulation hoses to the test system.   

 

2.3

 

Positioning the Unit 

Positioning the coolant circulator correctly is essential to its proper operation.  Firstly the unit should be 
positioned a reasonable distance away from the actual test system as vibration from the pump and fan 
will make mechanical measurements noisy.  If the test system is mounted on a vibration isolation table, 
DO NOT locate the coolant circulator on it as this will cause vibration noise to appear in the data. 

Since the head pressure generated by the pump is small, the 826A Water Cooler should be mounted 
slightly above or at about the same height as the test system.  Take care to mount the unit within ±0.5m 
of the height of the 800-series apparatus or the circulator will have trouble driving coolant through the 
entire loop. 

Содержание 826A

Страница 1: ...Aurora Scientific 2015 VER 1 00 page 1 Water Cooler 826A Publish Date 7 6 2015 Manual Version 1 00 www AuroraScientific com...

Страница 2: ...2015 VER 1 00 page 2 Copyright 2015 Aurora Scientific Aurora Scientific 25 Industry Street Aurora ON Canada L4G1X6 Tel 1 905 727 5161 Toll Free N America 1 877 878 4784 Fax 1 905 713 6882 Email techsu...

Страница 3: ...Water Cooler Instruction Manual Model 826A Aurora Scientific 2015 VER 1 00 page 3 Revision History Date Version Number Document Changes 07 06 2015 1 0 Initial Draft...

Страница 4: ...g the Liquid Reservoir 6 2 3 Positioning the Unit 7 2 4 Module Operation 8 3 Maintenance 9 4 Performance Guarantee Technical Support Warranty and Repair Information 10 4 1 Performance Guarantee 10 4 2...

Страница 5: ...e 5 1 Introduction The 826A Water Cooler is intended to be used as a simple liquid coolant circulator with Aurora Scientific 800 Series apparatus The Water Cooler incorporates a brushless DC pump to c...

Страница 6: ...e hose screws over the end of the hoses before mating the hose to the back of the circulator Once the hoses have been firmly pressed on tighten up the hose screws to secure the connection see Figure 1...

Страница 7: ...stopped Once full replace the cap and unplug the power adapter from the system until ready for use If the fluid refuses to flow through the system immediately when power is connected it is likely that...

Страница 8: ...y switching the 825A on or off Once the unit has been plugged in the fan speed can be set with the two position switch located on the back panel of the coolant circulator The low setting is in general...

Страница 9: ...t is full approximately 1 5cm or from the top and top up as necessary Aurora Scientific recommends replacing the coolant annually to keep your system operating optimally If you run out of coolant dist...

Страница 10: ...e Our performance guarantee states if for any reason a new product does not meet your research needs then you can return it to Aurora Scientific for exchange or a full refund The performance guarantee...

Страница 11: ...Scientific will be sent to the manufacturer for credit or replacement Ultimate disposition of these items will depend upon the manufacturer s decision All shortages must be reported within ten 10 day...

Страница 12: ...ntroller For muscle lever repairs we require the motor lever arm motor cable and control electronics Please don t send the power cord 3 Prepare Customs documents Canadian Clients no customs documents...

Страница 13: ...r choice or via parcel post If possible we prefer that you ship via FedEx You are responsible for both the shipping and brokerage charges so please mark the waybill accordingly Please don t ship freig...

Страница 14: ...cally commutated with brushless DC design Voltage Range 8 to 13 2 VDC Max System Pressure 22 PSI Head Pressure 390 m bar 12 VDC Maximum Flow 350 L h 1 54 US Gal min 12VDC Fan Voltage 12 VDC Current 0...

Отзывы: