Water Cooler Instruction Manual 

Model: 826A 

© Aurora Scientific, 2015                                        VER. 1.00 

page 8 

 

2.4

 

Module Operation 

Once all of the connections have been made and the coolant reservoir filled, the 826A is ready for use.  
The coolant circulator does not have a power switch, as it draws its power from a tap on the rear of the 
825A.  Circulation is therefore turned on by switching the 825A on or off.  Once the unit has been plugged 
in the fan speed can be set with the two position switch located on the back panel of the coolant 
circulator.  The low setting is in general sufficient to be used with all ASI 800 Series test systems for most 
temperature settings of the 825A.  Aurora Scientific recommends switching the fan setting to high when 

operating at either extremes of the TEC controller’s range.

 

The Module is equipped with dry break fittings to avoid any spillage of coolant in the event that the unit 
must be moved. 

 

 

Содержание 826A

Страница 1: ...Aurora Scientific 2015 VER 1 00 page 1 Water Cooler 826A Publish Date 7 6 2015 Manual Version 1 00 www AuroraScientific com...

Страница 2: ...2015 VER 1 00 page 2 Copyright 2015 Aurora Scientific Aurora Scientific 25 Industry Street Aurora ON Canada L4G1X6 Tel 1 905 727 5161 Toll Free N America 1 877 878 4784 Fax 1 905 713 6882 Email techsu...

Страница 3: ...Water Cooler Instruction Manual Model 826A Aurora Scientific 2015 VER 1 00 page 3 Revision History Date Version Number Document Changes 07 06 2015 1 0 Initial Draft...

Страница 4: ...g the Liquid Reservoir 6 2 3 Positioning the Unit 7 2 4 Module Operation 8 3 Maintenance 9 4 Performance Guarantee Technical Support Warranty and Repair Information 10 4 1 Performance Guarantee 10 4 2...

Страница 5: ...e 5 1 Introduction The 826A Water Cooler is intended to be used as a simple liquid coolant circulator with Aurora Scientific 800 Series apparatus The Water Cooler incorporates a brushless DC pump to c...

Страница 6: ...e hose screws over the end of the hoses before mating the hose to the back of the circulator Once the hoses have been firmly pressed on tighten up the hose screws to secure the connection see Figure 1...

Страница 7: ...stopped Once full replace the cap and unplug the power adapter from the system until ready for use If the fluid refuses to flow through the system immediately when power is connected it is likely that...

Страница 8: ...y switching the 825A on or off Once the unit has been plugged in the fan speed can be set with the two position switch located on the back panel of the coolant circulator The low setting is in general...

Страница 9: ...t is full approximately 1 5cm or from the top and top up as necessary Aurora Scientific recommends replacing the coolant annually to keep your system operating optimally If you run out of coolant dist...

Страница 10: ...e Our performance guarantee states if for any reason a new product does not meet your research needs then you can return it to Aurora Scientific for exchange or a full refund The performance guarantee...

Страница 11: ...Scientific will be sent to the manufacturer for credit or replacement Ultimate disposition of these items will depend upon the manufacturer s decision All shortages must be reported within ten 10 day...

Страница 12: ...ntroller For muscle lever repairs we require the motor lever arm motor cable and control electronics Please don t send the power cord 3 Prepare Customs documents Canadian Clients no customs documents...

Страница 13: ...r choice or via parcel post If possible we prefer that you ship via FedEx You are responsible for both the shipping and brokerage charges so please mark the waybill accordingly Please don t ship freig...

Страница 14: ...cally commutated with brushless DC design Voltage Range 8 to 13 2 VDC Max System Pressure 22 PSI Head Pressure 390 m bar 12 VDC Maximum Flow 350 L h 1 54 US Gal min 12VDC Fan Voltage 12 VDC Current 0...

Отзывы: