24
3. To turn alarm
limits on or off, tap on ʻʻON/OFFʼʼ button. (Turn on the
alarm before adjusting the value.)
NOTE: SpO
2
max limit is turned off by default.
NOTE: There is no alarm setting for PA (pulse amplitude).
4.
Tap on ʻʻ + ʼʼ or ʻʻ– ʼʼ buttons or drag the ʻʻseekbarʼʼ to adjust the values.
NOTE: The minimum alarm limit cannot exceed the max alarm limit,
even if the max alarm limit is turned off. For example, if the
max SpO
2
limit is turned off but was previously set at 90%,
the min SpO
2
limit cannot be set high than 90%. If you want
to set min SpO
2
limit at 90%, turn on the max SpO
2
limit, set
it above 90% and turn it off again as you wish.
NOTE: The following table describes the default settings,
adjustment ranges, and intervals.
High Priority Alarm
Factory
Default
Adjustment
Options
Adjustment
Interval
SpO
2
Upper Alarm Limit
Off
Off, 85 to 100
1% SpO
2
SpO
2
Lower Alarm Limit
85%
Off, 50 to 95
1% SpO
2
Pulse Rate Upper Alarm Limit
200 bpm
Off, 75 to 275
1 bpm
Pulse Rate Lower Alarm Limit
75 bpm
Off, 30 to 110
1 bpm
5.
Tap on ʻʻCONFIRMʼʼ to save the alarm limits.