Verifying System Operation
Use the following procedure to verify that the alarm system is working
properly.
1. Ensure the Infant Oximeter Module and Display Unit are both turned
on.
2. Ensure there is a connection established between the Infant Oximeter
Module
and Display Unit. (See “Device Pairing” section on page 16 for
more information.)
3. Place the Infant Oximeter Module on a flat surface.
4. Verify that an alarm message is displayed and that an alarm audio is
generated.
(See “Troubleshooting” section on page 31 if an alarm
message and audio signal is not generated.)
5.
Press on the “PAUSE AUDIO” to temporarily silence for 2 minutes.
6. After the alarm
signal is regenerated, press on the “AUDIO OFF” to
silence permanently the alarm signal.
NOTE: Alarm systems should be checked
before use.
Use the procedure below to monitor the readings (SpO
2
, pulse rate, pulse
amplitude) to verify that the device is functioning properly.
1. Set up Aulisa GA1001 system according to instructions above. (See
“Setting up Aulisa GA1001 system” section on page 13 for more
information.)
2. Ensure the Infant Oximeter Module is securely attached to the foot.
3. Verify the status indicator on the Infant Oximeter Module is blinking
green and that the Bluetooth connection status icon on the Display Unit
is blue.
4. Verify that a SpO
2
reading is displayed, that a pulse rate value
appears, and that a PA reading is displayed.