19 Speechmap Fitting Procedures
19.1 Section overview
FastFacts 19.1: Speechmap fitting procedures
19.2 Assessment data entry
FastFacts 19.2: Assessment data entry
Threshold data and parameters may be entered by scanning a barcode on a
previous Verifit or RM500SL printout or manually using a mouse, the keypad or
the numeric keypad on an external keyboard. See
2.7: Mouse, keyboard, barcode
scanner
, and
3.2: Input device operation.
To enter data from a barcode see
3.3: Barcode data input
and:
1)
Press the barcode trigger and center the red line on the barcode from a
distance of 6 - 8”. The scanner will beep and the Barcode Entry poster opens.
2)
Highlight and <PICK> [Done] to accept the
data indicated on the poster.
To enter assessment data manually:
1)
Highlight and <PICK> [Audiometry].
2)
Highlight and <PICK> an item to toggle its
setting or generate a list box. Age can be
changed only if Targets = [DSL child]. See
19.3: Assessment data choices - 1
and
19.4: Assessment data choices – 2
3)
Press <Continue> to proceed to data entry.
4)
Enter HL threshold using a mouse,keypad or keyboard. Left click the mouse on
the audiogram to enter a point. Click again to delete the point. On the keypad
the arrow keys change frequency and level and <PICK> enters or deletes a
threshold point. On a keyboard enter HL values on the numeric keypad.
5)
Press <Continue> when all points have been entered.
6)
Repeat the process for Bone conduction, UCL and RECD if [Enter] has been
selected for any. If RECD is set to [Measure], see
16: RECD (real-ear to
coupler difference)
.
The procedures in this section assume a familiarity with
10: Real-Ear
Measurement Setup
,
11: REM Screen Setup
and
18: Speechmap ®
.
This section covers the following topics:
19.2: Assessment data entry
19.7: Fitting to targets for mid-level speech
19.3:Assessment data choices - 1
19.8: Open fittings in Speechmap
19.4: Assessment data choices – 2
19.9: FM fitting and verification
19.5:Fitting to targets for soft speech
19.10: CROS fitting in Speechmap
19.6: Fitting to targets for loud sounds
Related sections are
18: Speechmap ®
and
20: Speechmap Technical Details