0611
Verifit User’s Guide Version 2.8
Page 48
9
Other Hearing Instrument Tests
9.1
Section overview
FastFacts 9.1: Other hearing instrument tests
9.2
Harmonic distortion
FastFacts 9.2: Harmonic distortion
In addition to the tests of ANSI S3.22, the Verifit provides the following:
Distortion
: Harmonic distortion at 2
nd
or 3
rd
harmonics and 2
nd
plus 3
rd
with input
levels varying from 60 – 90 dB SPL.
Directional
: A unique real-time broad-band test signal presented simultaneously
from left and right speakers and analysed into two separate reponse curves.
Multicurve
: 1 – 4 curves of coupler SPL or coupler gain for swept tones or pink
noise in 1/12
th
octaves at selected input SPL of 0 or 40 to 90 dB in 5 dB steps.
Can be used as a 1/12th octave
spectrum analyzer
by setting stimulus to 0 dB.
Battery drain
: A 10 second test of battery current drain with no acoustic input, a 1
kHz 65 dB input SPL, and an average for 90 dB input SPL at the HFA or SPA
frequencies selected in Setup.
Manual control
: Coupler and HIT reference mic. SPL at selected 1/12
th
octave
frequencies and stimulus SPL of 0 and 40 – 90 dB in 5 dB steps. Can be used as
a dual channel
sound level meter
with 1/12
th
octave, A and C filters.
This section covers the following topics:
9.2: Harmonic distortion
9.8: Multicurve results
9.3: Directional function test
9.9: Spectral analysis in Multicurve
9.4: Directional ITE positioning
9.10: Battery drain
9.5: Directional BTE positioning
9.11: Manual control procedure
9.6: Directional HIT procedure
9.12: Sound level meter using manual control
9.7: Multicurve procedure
This test measures harmonic distortion at 2
nd
or 3
rd
harmonics with input levels
varying from 60 – 90 dB SPL. It also presents the sum of 2
nd
and 3
rd
harmonics as
total harmonic distortion (THD). For hearing aids, this differs little from true THD.
1)
Press <Tests>, then highlight and <PICK> [Distortion].
2)
Refer to
7.11: Coupling the hearing instrument
,
7.12: Positioning the hearing
instrument
and
2.6: Battery pill use and care
to prepare the instrument.
3) Turn the hearing instrument ON
.
4)
Highlight and <Pick> [Start test].
5)
Highlight and <PICK> [Harmonics], then highlight and <PICK> THD, 2nd or 3
rd
from the drop-down list.
6)
Highlight and <PICK> [Input], then highlight and <PICK> the desired input level
(60 – 90 dB SPL) from the drop-down list.
7)
Press <Continue> to run the test. Distortion will be displayed at 1/3 octaves in
graphic and in tabular form.