Verifit User’s Guide Version 2.8
Page 41
8.4
ANSI 1996 Linear and AGC tests
FastFacts 8.4: ANSI 1996 Linear and AGC tests
1)
Press <Tests>, then highlight and <PICK> [Linear] or [AGC].
2)
Highlight and <PICK> [Start test]; test setup instructions will be displayed.
3)
Refer to
7.11: Coupling the hearing instrument
,
7.12: Positioning the hearing
instrument
and
2.6: Battery pill use and care
to prepare the instrument for
testing.
4) Turn the instrument ON
.
5)
Set the gain control (manual or programmable) to maximum with all other
controls set to provide the widest frequency response, greatest gain and
highest maximum output. AGC controls and/or programmable parameters
should be set as designated by the manufacturer.
6)
Press <Continue> to run the test or press <Cancel> to abort the test. Note
that, if required, the Verifit will pause during the test to request that the volume
control be adjusted to the Reference Test Position (RTP). Open the HIT
chamber if necessary and, without moving the hearing instrument, adjust the
user gain control until the large vertical arrow on the screen prompt
disappears. Press <Continue> to resume the test.
If you are unsure which test sequence to run, try [AGC] first. If attack and release
times are all near zero, the instrument is probably linear.