MAN - BBR2S - REV: Original | E. Operation
37
The diagnostics button will not be shown on the main menu unless the currently logged in user has permission for
this screen.
Analog Inputs AI1, AI2, and AI3
Raw voltages into the system. You should be able to use a voltmeter to test the voltage at the input pin and it should
match what is shown here. AI1 is deflection, AI2 is load, and AI3 is temperature.
Analog Outputs AO1 and AO2
Raw voltages out of the system. Use a voltmeter to test the voltage at the output pin and verify that it matches the
value shown on the diagnostics screen.
Digital Inputs 1, 2, 3, and 4
Raw values into the system. Use a voltmeter to test the state at the input pin and verify that it matches the value
shown on the diagnostics screen. DI1 and DI2 are used to bring in the auto/zero/load switch. DI3 and DI4 are not
used at this time.
Digital Outputs 1 and 2
Raw values out of the system. Use a voltmeter to test the state at the output pin and verify that it matches the value
shown on the diagnostics screen. DO1 is the valve to switch between contact and test loads. DO2 is not used at this
time.
Digital Outputs 3 and 4
High speed pulse outputs. Use a scope to test the output pin and verify that it matches the frequency and duty cycle
entered in the diagnostics screen. DO3 and DO4 are not used at this time.
Figure E.14 - Diagnostics main software screen