MAN - BBR2S - REV: Original | E. Operation
35
4. Use the mode switch to move the load control switch to the load position.
5. Adjust the load regulator until the load display reading is 980 ± 50 mN.
6. Use the mode switch to move the load control switch between zero (contact load) and load (test load) several
times to recheck the specimen load and verify that the values are stable.
NOTE:
Allow the specimen to stabilize in the cooling fluid bath at the required test temperature
according to the specifications.
7. When finished, place the mode switch in the AUTO position.
Start Test
1. Raise the loading shaft. Remove the non-compliant beam and place the specimen in the test position.
2. Gently lower the loading shaft, so it contacts the specimen. Make sure the load display is 35 ±10 mN. If it is not,
adjust the zero regulator.
3. Press the START TEST button.
4. The horizontal graph allows the operator to monitor test progress. Current deflection and load values are
displayed on the graph.
5. To abort the test, press the STOP TEST button.
6. When the test is complete the system will automatically generate a report and display it.
7. If you have connected a printer to your BBR2S, you can print the test summary report from this screen. To do so,
select file and then print.
Run Crack Sealant Test
1. Before running this test you will need to remove the (2) anvil adapters from the Anvil. This is performed by lifting
STRAIGHT UP on each adapter off the guide pins.
WARNING: Load frame should be removed from the bath and allowed to warm to room
temperature before removing the anvil adapters.
2. Raise the loading shaft. Remove the non-compliant beam and place the specimen in the test position. Gently
lower the loading shaft, so it contacts the specimen. Make sure the load display is 35 ±10 mN. If it is not, adjust
the zero regulator.
3. Press the START TEST button.
4. The horizontal graph allows the operator to monitor test progress. Current deflection and load values are
displayed on the graph.
5. To abort the test, press the STOP TEST button.