33
Reprocessing
5.3.4 Wipe disinfection
1. Disinfect the surface evenly with a suitable cloth and suitable disinfectant. Pay
particular attention to hard-to-reach areas.
2. Wait for the contact time to elapse.
5.4 Reprocessing the accessories
5.4.1 Overview
Accessory
Disposable pr
oduct
Max. r
epr
ocessing
cycles
After each use
After each patient
Daily
W
eekly
Every 14 days
Monthly
Pr
e-tr
eatment
Pr
e-cleaning
Manual cleaning and disinfection
Mechanical cleaning and disinfection
Sterilisation
Secretion canister system
Secretion canister
50
X
X
X
X
X
Secretion canister lid
50
X
X
X
X
X
Double hose connector
50
X
X
X
X
X
Bacterial filter
1
X
Hoses
Suction hose
60
X
X
X
X
X
Vacuum hose
60
X
X
X
X
X
Connecting hose
60
X
X
X
X
X
1
Immediate filter change in the event of discolouration, contamination, oversuction.
The bacterial filter is no longer in optimum condition if the vacuum displayed is above
−0.3 bar when the vacuum regulator is in the ”MAX” position and the suction hose is
open.
5.4.2 Selecting process chemicals
Follow the manufacturer’s instructions for the process chemical.
Cleaning agent
(manufacturer)
Active ingredients in 100 g
Type
Secre-
tion
canister
system Hoses
Disinfectants – Manual reprocessing
Gigasept® FF neu (Schülke
& Mayr)
<5% phosphonates, <5% anionic surfactants, <5%
non-ionic surfactants, perfumes, methylisothiazoli-
nones
Liquid
concentrate
X
Cleaning agent
(manufacturer)
Active ingredients in 100 g
Type
Secre-
tion
canister
system Hoses
Cleaning agents – Mechanical reprocessing
neodisher® MediClean forte
(Dr. Weigert)
<5% non-ionic and anionic surfactants, enzymes
Liquid
concentrate
X