35
Reprocessing
Checking and
maintaining
1. Check whether reprocessing was successful with a
suitable light magnifier.
• Free of particles and organic material
2. Dispose of damaged accessories or have them repaired.
Assembly
Not necessary.
Function check
Not necessary.
Packaging
1. Label the accessories.
2. Pack the accessories using a packaging system according
to DIN EN ISO 11607.
Sterilisation
Prefractionated
vacuum:
3x
Temperature:
134 °C / 273 °F
Time:
5 min
Dry:
10 min
1. Sterilise the accessories using a suitable procedure:
• Steam sterilisation / autoclaving
Ideally, always use the same procedure.
Steriliser:
• in accordance with EN 285
Storage
1. Observe the ambient conditions; see chapter „11 Techni-
5.4.4 Hoses
Pre-treating at the site
of use
Flushing:
5x / 30 s
1. Clean the hoses under cold, running water.
2. Rinse the hoses thoroughly.
» No more residue is visible.
Collecting and
transporting
1. Label damaged hoses.
2. Place the hoses in a canister.
3. Close the canister.
4. Transport the canister to the reprocessing site.
Pre-cleaning
Flushing:
5x / 30 s
Pre-cleaning is only necessary for mechanical cleaning
and disinfection.
1. Clean the hoses evenly under running water.
2. Rinse the hoses thoroughly under running water.
Disassembly
Not necessary.