Reprocessing
Unsuitable cleaning agents and disinfectants.
The product can become damaged.
• Do not use any process chemicals containing the following ingredients
on plastic
parts
:
• Chloramides or phenol derivatives
• Do not use abrasives.
Incorrect mechanical cleaning and disinfection.
Corrosion due to moisture.
•
Remove the products immediately after the programme is finished.
5.2 Preparing and completing reprocessing
Prior to reprocessing
1. Disassemble the product into the following items for reprocessing:
• Secretion canister system (secretion canister, secretion canister lid, double hose
connector, bacterial filter)
• Hoses (suction hose, vacuum hose, connecting hose)
After reprocessing
1. Perform a function check.
5.3 Preparing surfaces
5.3.1 Overview
Surface
After each use
After each patient
Daily
W
eekly
Every 14 days
Monthly
Pr
e-cleaning
Wipe cleaning
Wipe disinfection
Spr
ay disinfection
Remarks
Coated surfaces
X
X X
As per agent manufacturer’s instructions
Other surfaces
X
X X
As per agent manufacturer’s instructions