28 Operation
Display
Cause
Troubleshooting
The secretion canister is
full.
Change the secretion
canister.
In the event of strong
foaming:
Fit the anti-foaming device
over the sensor.
The device temperature is
too high.
Please provide sufficient
ventilation.
Service required!
Device cannot be operated. Please
contact service
Service is required.
Contact ATMOS Service.
4.12 Trolley
1. Secure the device to the trolley tray
using the two screws.
» The
Soll - 500 mbar
- 250
Ist
mbar
AUTO
symbol appears in the
display.
2. Attach the 3-litre secretion canister to
the trolley.
4.13 Foot controller
The vacuum can be set by connecting the foot controller:
1. Plug the foot controller into its connection before switching on the device.
2. Connect the suction catheter, suction attachment or suction instruments to the
suction hose.
3. Switch on the ATMOS S 351 OT. Make sure that the pilot lamp in the ON / OFF switch
lights up.
4.
Select the final vacuum required using one of the buttons (5), (6) and (7). Keeping the
– (5) and + (6) buttons pressed changes the value faster.
5. You can now vary the vacuum between 0 and the maximum value set using the foot
controller.