Maintenance and service
6 Maintenance and service
Maintenance, repairs and periodic tests may only be carried out by persons who have
the appropriate technical knowledge and who are familiar with the product. The per-
son must possess the test devices and original spare parts required to carry out these
measures.
ATMOS recommends: Work should be carried out by an authorised ATMOS service
partner. This ensures that repairs and tests are carried out professionally, original spare
parts are used and warranty claims remain unaffected.
Maintenance, repairs and periodic tests may
not
be carried out while the product is
being used on the patient.
6.1 Periodic tests
Perform a repeat test of the electrical safety according to IEC 62353 at least every 12
months.
In this context, ATMOS recommends conducting an inspection in accordance with the
manufacturer‘s specifications.
6.2 Sending in the device
1. Remove all consumables and dispose of them properly.
2. Clean and disinfect the product and accessories according to the operating instruc-
tions.
3. Enclose any used accessories with the product.
4. Fill in form QD 434 ‘Delivery complaint / return shipment’ and the corresponding
Decontamination certificate
.
This form is enclosed with each delivery and can be found at www.atmosmed.com.
5. The product must be well padded and packed in suitable packaging.
6. Place form QD 434 ‘Delivery complaint / return shipment’ and the corresponding
Decontamination certificate
in an envelope.
7.
Affix the envelope to the outside of the package.
8. Send the product to ATMOS or your dealer.