Notes on EMC
Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS
®
E 341 Battery is intended for use in the electromagnetic environment specified below. The
customer or user of the ATMOS
®
E 341 Battery should ensure that it is used in such an environment.
Immunity Test
IEC 60601- Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Conducted RF IEC
61000-4-6
3 V
eff
150 kHz to 80 MHz
10 V
Portable and mobile communications
equipment should be separated from the
ATMOS
®
E 341 Battery incl. the cables by
no less than the distances calculated/listed
below.
Recommended distances:
d = 0.35
√
P
d = 0.35
√
P
80 MHz to 800 MHz
d = 0.70
√
P
800 MHz to 2.5 GHz
where „P“ is the max. power in watts (W) and
D is the recommended separation distance in
meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compliance
level (b).
Interference may occur in the vicinity of
equipment containing following symbol:
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
10 V/m
NOTE 1
With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by
absorption and reflection of buildings, objects and people.
a
The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain
radio equipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined
exactly.
To determine the electromagnetic environment in regard to stationary transmitters, a study of the location
is to be considered. If the measured field strength at the location where the ATMOS
®
E 341 Battery is used
exceeds the above compliance level, the ATMOS
®
E 341 Battery is to be observed to verify the intended
use. If abnormal performance characteristics are noted, additional measures might be necessary, e. g. a
changed arrangement or another location for the device.
b
Within the frequency range of 150 kHz to 80 MHz the field strength should be below 3 V/m.
The device may not be used directly next to other devices or piled up with other devices. If operation next
to or piled with other devices is necessary, please watch the device to check its intended operation in this
arrangement.
Recommended separations between portable and mobile RF
Communications equipment and the ATMOS
®
E 341 Battery
The ATMOS
®
E 341 Battery is intended for use in electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the ATMOS
®
E 341 Battery can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
Communications equipment and the ATMOS
®
E 341 Battery as recommended below, according to the
maximum output power of the communications equipment.