40
Cleaning and disinfection
5.5 Hygienic plan
What
How
Notices
When
Who
Parts to be reprocessed
C D S
Daily
*
Every 14 days
*
After each patient / after each suction
Qualifi ed and trained staff who
are familiar with reprocessing
Cleaning
Disinfection
Sterilization
Surfaces
Housing
X
X
Wipe cleaning and disinfection
X
X
Device base
X
X
1
Wipe cleaning and disinfection
X
X
Hose reel
X
X
1
Wipe cleaning and disinfection
X
X
Wall and device support
X
X
Wipe cleaning and disinfection
X
Power supply and recharging
unit
X
X
Wipe cleaning and disinfection with a damp
cloth. Do not immerse into any liquid!
X
Secretion canister system
DDS secretion canister
X
X
1
X
2
Cleaning with a brush; cleaning and
disinfection (automatic or manual),
sterilization is possible
X
X
Outer canister lid (DDS)
X
X
1
X
2
Cleaning with a brush; cleaning and
disinfection (automatic or manual),
sterilization is possible
X
X
Inner canister lid (DDS)
X
X
1
X
2
Cleaning with a brush; cleaning and
disinfection (automatic or manual),
sterilization is possible
X
X
Float ball (DDS)
X
X
1
X
2
Cleaning with a brush; cleaning and
disinfection (automatic or manual),
sterilization is possible
X
X
Filter holder
X
X
1
X
2
Cleaning with a brush; cleaning and
disinfection (automatic or manual),
sterilization is possible
X
X
Sealing ring
X
X
1
Cleaning and Disinfection
X
X
Bacterial fi lter
Exchange. If the fi lter is blocked it must also
be exchanged.
X
X
Fingertip
Exchange.
X
X
Support for canister system
X
X
1
Cleaning and Disinfection
X
X
Suction hose (DDS)
X
X
1
X
2
When using without suction catheter
X
X
X
X
1
X
2
When using with suction catheter
X
X
Vacuum hose
X
X
Cleaning and Disinfection
X
X
Cleaning and disinfection plan
ATMOS
®
C / E 341 Battery
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16
79853 Lenzkirch/Germany
Phone +49 7653 689-0
Fax +49 7653 689-190
www.atmosmed.com
GA3GB.320100.0 2017-11 Index: 02
Recommended disinfectants
Surface disinfection for coated surfaces:
•
Green & Clean SK (ATMOS)
• Dismozon
®
pur (Bode Chemie)
• Dismozon
®
plus (Bode Chemie)
•
Kohrsolin
®
FF (Bode Chemie)
•
Kohrsolin
®
extra (Bode Chemie)
• Perform
®
(Schülke & Mayr)
•
Terralin
®
Protect (Schülke & Mayr)
Other surfaces:
• Dismozon
®
pur (Bode Chemie)
• Dismozon
®
plus (Bode Chemie)
•
Kohrsolin
®
FF (Bode Chemie)
•
Kohrsolin
®
extra (Bode Chemie)
•
Mikrobac
®
forte (Bode Chemie)
• Perform
®
(Schülke & Mayr)
•
Terralin
®
Protect (Schülke & Mayr)
• SaniCloth
®
Active (Ecolab)
• Incidin
®
Active (Ecolab)
•
Mikrozid
®
Sensitive Wipes (Schülke
& Mayr)
Manual disinfection of instruments:
•
Mucocit
®
-T (Merz Dental)
•
Gigasept
®
FF neu (Schülke & Mayr)
•
Sekusept
®
PLUS (Ecolab)
For concentrations, contact time, tem-
perature, material compatibility, please
see the relevant information from the
manufacturer.
Important information
Wipe cleaning and disinfection:
All surfaces have to be wiped with
a clean (disposable) wipe which is
damped with disinfectant solution.
The entire surface has to be wiped
thoroughly and may not be dried
afterwards.
1)
Preferred: mechanical cleaning and
disinfection in the washer disinfector
with a device according to the ISO
15883-1 (Program: Rinsing 1 min with
cold water, cleaning 5 min at 55°C,
neutralisation 1 min with 1/3 cold and
2/3 warm water, rinsing 1 min with
purifi ed water, thermal disinfection 5
min at 93°C with purifi ed water)
2)
If required hot-steam sterilization at
134°C, 3 x fractionated pre-vacuum
method, sterilization time 5 min with a
device according to EN285
* Homecare, unless there is no change
in patient.
Wrong concentration of
disinfectants may lead to
material
damage!
The above stated hygiene requirements are based on
the regulations according to the Medical Devices Act, the
Medical Devices Operator Ordinance, §18 IfSG and the
recommendations of the Robert Koch Institute.
Defi nition of the required reprocessing steps result
from the recommendations of the Robert Koch Institute:
„Requirements for the reprocessing of medical products“,
from Robert Koch Institute. The medical products were
categorised in the risk groups uncritical, semi-critical and
critical. The reprocessing measures mentioned in this
cleaning and disinfection plan are a recommendation
of ATMOS MedizinTechnik. Any additional reprocessing
measures are at the operator‘s discretion.
All the recommended disinfectants which are stated
herein are listed disinfectants (VAH/RKI) and have been
tested on their suitability of use on the ATMOS
®
C / E
341 Battery. ATMOS MedizinTechnik cannot be hold
liable for any damage caused by wrong concentration
of the disinfectants or by the application of any other
disinfectants.
Patients with suspicion of a clinical disease or who
developed a transmissible spongiform encephalopathy
(CJK, vCJK, etc.) have to be treated at facilities which are
able to provide for the necessary preventive measures
against infection. The reprocessing of the reusable
instruments and material may only be performed
at facilities which have an externally certifi ed QM
Management acc. to DIN EN ISO 13485.
The Medical Devices Act, IfSG, the RKI directives, BGR 250
and TRBA 250 always have to be considered.