54
Technical data
11.0 Technical data
Device
Dimensions (W x H x D):
with DDS canister system
with Serres
®
canister system
with Medi-Vac
®
canister system
with universal bracket
370 x 277 x 146 mm
370 x 277 x 146 mm
370 x 277 x 136 mm
370 x 277 x 136 mm
Weight:
Device with battery / without canister
system and support
DDS canister system with support for
DDS canister system
Serres
®
canister system with support for
Serres
®
canister system
Medi-Vac
®
canister system with support
for Medi-Vac
®
canister system
Universal bracket
3.65 kg
1.00 kg
0.65 kg
0.295 kg
0.2 kg
Operation:
Temperature range
Relative air humidity
Air pressure
-5° C up to +50° C
5 % to 95 % without condensation
540 hPa to 1100 hPa
Transport / storage:
Temperature range
Relative air humidity
Air pressure
-40° C up to 70° C
5 % to 95 % without condensation
540 hPa to 1100 hPa
Charging:
Temperature range
Relative air humidity
Air pressure
0° C up to 40° C
5 % to 95 % without condensation
540 hPa to 1100 hPa
Maximum operating altitude
5000 m (NN)
Contamination level
Class 1 (fully-sealed housing)
Overvoltage category
II
Maximum power consumption
45 W
Maximum current consumption
3.8 A
Mains voltage
12 V DC nominal (at least 10 V, max. 15 V)
at the charging interface or via the power
supply and recharging unit.
Pump
Vacuum pump (diaphragm pump), 1 head
Suction capacity at the device inlet
(without canister system) at -0.8 bar, fully
recharged battery and 21° C / 1013 hPa
(determined by buffer canister 1 l)
34 l/min ± 4 l/min
Suction capacity at the inlet of the
DDS canister system at -0.8 bar, fully
recharged battery and 21° C / 1013 hPa
30 l/min ± 3 l/min
Maximum achievable vacuum
0.8 bar* ± 5 % resp. 80 % of the air
pressure