47
Eliminating errors
7.0 Eliminating errors
The ATMOS
®
E 341 Battery was subjected to a thorough quality control in the factory.
Nevertheless, if a problem may occur you can possibly solve it yourself.
Recharging and battery status
Error indication
Possible cause
Remedy
Device cannot be
recharged.
The plug from the
charging accessories is
poorly fitted.
1. Check the connection to the
mains supply.
Charging accessories
are defective.
1. Exchange charging accessories.
Battery is not
connected properly.
1. Check the plug connections in
the battery compartment.
Battery temperature
too high or too low.
1. After long-term use: Let the
device cool down.
2. Extreme ambient temperature:
Position the unit where
appropriate to a cooler or
warmer place.
Deep discharge of the
battery.
1. Exchange the battery.
Defective electronics. 1. Send in the device for repair.
LED on the power supply
and recharging unit is not
illuminated.
Defective power
supply and recharging
unit.
1. Exchange the power supply
and recharging unit.
The mains plug is
poorly fitted.
1. Check the connection to the
mains supply.
When recharging the
battery 100 % cannot be
achieved. The charging
time can take up to 3
hours.
Battery service life
is exhausted or the
battery is defective.
1. Exchange the battery.
Wrong charging
accessories.
1. Only use the provided charging
accessories or an original spare
part.
The bottom green and
the red LED on the
display of battery status
flashes and a signal tone
sounds every 5 s.
Battery is almost
completely
discharged.
1. Recharge the battery.
During switch on: All
LEDs of the display of
battery status flash for 5
s, a signal tone sounds.
Remaining battery
level is low.
1. Perform a function check.
2. Exchange the battery.
All the green LEDs on the
display of battery status
flash permanently.
A non-ATMOS battery
is used.
1. Only use the provided battery
or an original spare part.