33
Cleaning and disinfection
Porous plastic due to detergents containing solvents.
1. Dampen a soft cloth with water and a mild, solvent-free detergent.
Damaged headphone capsules due to water.
2. Make sure that the headphone capsule does not come into contact with water.
3. Clean the following items with a damp cloth:
•
Headphone bow.
•
Headphone ear cushions.
•
Bone vibrator capsule.
•
Patient response button.
5.2 Overview hygiene measures
Device part
Cleaning
Disinfection
Recommendation
After each
patient
Device
x x mild solvent-free detergent
x
Headphone bow
x
mild solvent-free detergent
x
Ear cushions
x
mild solvent-free detergent
x
Bone vibrator capsule x
mild solvent-free detergent
x
Patient response
button
x
mild solvent-free detergent
x
5.3 Recommended disinfectants
If you are using aldehyde and amine-containing disinfectants on the same object, this
may cause discolouration.
Housing damage due to unsuitable detergents.
Corrosion damage and stress cracks are possible.
•
Only use the recommended disinfectant or disinfectant solutions with the same
ingredients.
•
Do not use disinfectants containing alcohol.
•
Do not use disinfectants containing organic and inorganic acids or bases.
•
Do not use disinfectants containing chloramides or phenol derivatives.
Disinfectant
Ingredients
in 100 g Manufacturer
Green & Clean SK
Di alkyl dimethyl ammonium chloride
< 1 g Metasys, Rum
(Austria)
Alkyl dimethyl ethyl benzyl ammonium
chloride
< 1 g
Alkyl dimethyl benzyl ammonium chloride
< 1 g
Dismozon
®
plus
(Granulate)
magnesium monoperoxyphthalate hexahy-
drate
95.8 g Bode Chemie,
Hamburg
Kohrsolin
®
FF
(Application concen-
trate)
glutaral
5 g
Bode Chemie,
Hamburg
benzyl-C12-C18-alkyldimethyl-ammonium
chlorides
3 g
Didecyldimethylammonium chloride
3 g
ATTENTION