19
ENGLISH
3.3.2 Setting the vacuum
1. Switch on the device using the push-
button (fig. 1(F)).
2. Keep the opening of the suction tube
(fig. 15) and the fingertip (fig. 16) shut.
3. Turn the vacuum regulator (fig. 1(E))
until the vacuum meter (fig. 1(D))
displays the desired vacuum. Please
note that higher vacuum values are
set by turning the vacuum regulator
clockwise and that the indicator of the
manometer moves counterclockwise
at the same time.
4. You can also perform fine adjustments
to the set vacuum via the additional air
opening on the fingertip.
5. To switch off the device, hold the push-
button (fig. 1(F)) pressed for several
seconds.
Fig. 15
Fig. 16
3.3.3 Aspiration
1. Perform the aspiration procedure as you
were instructed by specialized personnel.
2. Control the aspiration procedure and the
desired vacuum via the additional air ope-
ning on the fingertip (example in fig. 16).
3. Decommission the device after the aspira-
tion procedure according to chapter 3.4.
Please note that the start-up of the
pump is not possible from an existing
vacuum of – 0.5 bar when switching
on again.
In any case, please take note of further
important information in section 2.3.1
regarding the handling of the ASSKEA
double filter system!
During the aspiration procedure, en-
sure that the battery has a sufficient
capacity. The display of the battery
condition must signalize green or
yellow. If the display turns red, the
external power supply unit must
be plugged in immediately. In this
case, the aspiration can be continued
without any interruption.