3.2.1 Connecting the ASSKEA reusable secretion canister system
Fig. 3
Fig. 4
Fig. 5
1. Screw the lid of the reusable secretion canister sufficiently tight and attach the
canister to the device.
2. Insert the connecting tube with the blue elbow fitting
toward the side of the device. Make sure that the side of
the external bacterial filter with the lettering ‘IN’ is facing
the canister (directed toward patient)!
3. Place the grey elbow adapter of the connection tube onto
the connection in the middle of the secretion canister lid
(above the overflow protection).
4. Place the suction tube onto the second connection
(patient connection). Ensure that the tube is fixed
sufficiently tight.
needed. Make sure that the device is positi-
oned in such a way that it is possible to easily
disconnect it later.
Information on the permissible environmen-
tal conditions during operation can be found
in chapter 7 ‘Technical data’.
3.2 Startup
• Remove the device and the accessories
from the packaging.
• Always place the device on a sturdy and flat
surface.
• Prior to the initial startup it is essential that
you follow the safety instructions in chapter
1.6.
• Always have one backup external bacterial
filter ready, since it is absolutely necessary
for safe operation (does not apply for dis-
posable secretion canister system).