30
ENGLISH
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ASSKEA secretion aspirators MPlus-series are intended for operation in the electromagnetic environment specified below.
The customer or user of an ASSKEA secretion aspirator MPlus-series must ensure that it is operated in such an environment.
Immunity testing
IEC 60601 test level
Compliance level
Electromagnetic Environment – Guidance
Conducted RF disturbances
pursuant to IEC 61000-4-6
Radiated RF disturbances
pursuant to IEC 61000-4-3
3 V
effective value
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5GHz
3 V
3 V/m
Portable and mobile RF communications equipment should be
used no closer to the ASSKEA secretion aspirator MPlus-series
incl. cables than the recommended separation distance calcula-
ted from the equation applicable to the frequency of the trans-
mitter.
Recommended separation distance:
d= 1.2
√
P
d= 1.2
√
P for 80 MHz to 800 MHz
d= 2.3
√
P for 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the data provided by the transmitter
manufacturer and d is the recommended separation distance in
meters (m).
The field strength from fixed RF transmitters, as determined by
an electromagnetic site survey
a
, should be lower than the com-
pliance level
b
in each frequency range.
Interference is possible in the vincinity of devices bearing the
symbol shown below.
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Содержание M20 Plus
Страница 1: ...M20Plus M30Plus Gebrauchsanweisung Instruction Manual ...
Страница 34: ......
Страница 35: ......