Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ASSKEA secretion aspirators MPlus-series are intended for operation in the electromagnetic environment specified below.
The customer or user of an ASSKEA secretion aspirator MPlus-series must ensure that it is operated in such an environment.
Immunity testing
IEC 60601 test level
Compliance level
Electromagnetic Environment – Guidance
Electrostatic discharge (ESD)
pursuant to IEC 61000-4-2
±6kV contact discharge
±8kV air discharge
±6kV contact discharge
±8kV air discharge
Floors should be made of wood, concrete or ceramic tiles. If
the floor is covered with synthetic material, the relative humidity
should be at least 30 %.
Fast transient electrical
disturbances/bursts pursuant
to IEC 61000-4-4
±2kV for power supply lines
±1kV for input and output lines
±2kV for power supply lines
±1kV for input and output
lines
Main supply voltage should be of the same quality as in a typical
commercial, living or hospital environment.
Interference voltages/Surges
pursuant to IEC 61000-4-5
±1kV voltage differential mode
±2kV common mode
±1kV voltage differential
mode
±2kV common mode
Main supply voltage should be of the same quality as in a typical
commercial, living or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
pursuant to IEC 61000-4-11
<5% U
T
(>95% dip
of U
T
) for ½ period
40% U
T
(60% dip
of U
T
) for 5 periods
70% U
T
(30% dip
of U
T
) for 25 periods
<5% U
T
(>95% dip
of U
T
) for 5s
<5% U
T
(>95% dip
of U
T
) for ½ period
40% U
T
(60% dip
of U
T
) for 5 periods
70% U
T
(30% dip
of U
T
) for 25 periods
<5% U
T
(>95% dip
of U
T
) for 5s
Main supply voltage should be of the same quality as in a
typical commercial, living or hospital environment. If the user
of an ASSKEA secretion aspirator MPlus-series requires
continued operation also during mains power outages, it is
recommended to power the ASSKEA secretion aspirator MPlus-
series from an uninterruptible power supply or a battery.
Magnetic field at power
frequency (50/60 Hz) pursuant
to IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteris-
tic for a typical location in a typical commercial, living or hospital
environment.
Note: U
T
is the AC mains voltage prior to the application of the test levels.