SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-1
7
DIAGNOSTICS AND
TROUBLESHOOTING
INTRODUCTION
This section explains:
•
General troubleshooting using built in diagnostic tools
•
BIS VIEW Monitoring System troubleshooting procedure.
•
Status messages, possible causes, and corrective actions
7.1
General Troubleshooting
The BIS VIEW Monitoring System has both automatic and manual diagnostic features.
These features check the BIS VIEW Monitoring System’s operability and status, and report
software and hardware malfunctions.
IF ANY FAILURES ARE NOTED, SEE SECTION 7.3 “BIS
VIEW System Messages and Corrective Actions”
The
power-up
diagnostics run automatically the first time the unit is powered on, after the
battery or power supply have been replaced, or after the Reset button has been pressed.
Software image checksums and trend data memory are tested and repaired if necessary.
The
automatic
diagnostics run continuously in the background while the unit operates.
These procedures check the BISx for the following conditions: interface faults, disconnect,
lead off, and power faults.
The
manual
diagnostics are operator initiated using the monitor soft keys and menu
choices. These procedures check the digital signal converter function within the BISx.
NOTE:
The manual diagnostics can be run safely while the patient is connected to
the BIS VIEW Monitoring System; however, running the diagnostics will
temporarily disrupt monitoring. Do not run the DSC Self Test during
electro-cautery as it may erroneously indicate a failure of the BISx.