SECTION 4 PREPARATION FOR USE AND INSTALLATION
_______________________________________________________________________
4-5
•
Verify that light to right of ON/Standby button is green.
•
Verify all self-tests complete successfully. A beep tone sounds.
•
Verify next screen says “Connect BISx.”
4.
Connect BISx to monitor. Connect PIC and sensor to BISx.
•
Verify screen says, “BISx Initialization Complete.”
•
Verify SENSOR CHECK begins.
5.
Disconnect power cord from rear of monitor.
•
Verify ‘OPERATING ON BATTERY BACKUP’ is displayed.
•
Verify battery icon displays below BIS number.
6.
Reconnect power cord.
•
Verify battery icon is not displayed below BIS number.
•
Verify “OPERATING ON BATTERY BACKUP” is not displayed.
7.
End of install.