SECTION 1 SAFETY PRECAUTIONS
________________________________________________________________________
1-3
DUE TO ELEVATED SURFACE TEMPERATURE, DO NOT PLACE BISx IN
PROLONGED DIRECT CONTACT WITH PATIENT’S SKIN, AS IT MAY
CAUSE DISCOMFORT.
UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT
CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS
MATERIALS. PLACE CONTAMINATED MATERIALS IN REGULATED
WASTE CONTAINER.
DO NOT MIX DISINFECTING SOLUTIONS (e.g., BLEACH AND AMMONIA),
AS HAZARDOUS GASES MAY RESULT.
WHEN CONNECTING EXTERNAL EQUIPMENT (e.g., DATA CAPTURE
COMPUTER), THE SYSTEM LEAKAGE CURRENT MUST BE CHECKED
AND MUST BE LESS THAN THE IEC601-1-1 LIMIT.
THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE
EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY
LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM.
CONSIDERATION RELATING TO THE CHOICE SHALL INCLUDE:
•
USE OF THE ACCESSORY IN THE PATIENT VICINITY
•
EVIDENCE THAT THE SAFETY CERTIFICATION OF THE
ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO THE
APPROPRIATE IEC 60601-1 AND/OR IEC 60601-2-26 HARMONIZED
NATIONAL STANDARD.
POWER SUPPLY IS INTERNALLY FUSED. REPLACE POWER SUPPLY ONLY
WITH ASPECT MEDICAL SYSTEMS BIS VIEW POWER SUPPLY.
1.2
Cautions
Read this entire manual carefully before using the monitor in a clinical setting.
To completely remove power from the unit: put the monitor in standby mode,
disconnect power cord from the power cord receptacle of the monitor, then remove
the battery from the monitor.
Continuous impedance checking may need to be disabled if the 1 nanoampere 128
Hz impedance check signal interferes with other equipment (e.g., evoked potential
monitors).
Do not autoclave the BISx or Monitor. Autoclaving will seriously damage both
components.