SECTION 6 PREVENTIVE MAINTENANCE
_______________________________________________________________________
6-8
Safety
Tester
measured by connecting the patient connection terminals of a safety tester to a Sensor
Simulator that is connected to the PIC.
To check BISx patient isolation:
1.
Connect the Sensor Simulator to the PIC in place of a sensor. (See Appendix.)
2.
Use a jumper to short the two terminals at the end of the simulator .
3.
Connect the test lead to the shorted terminals. Make sure that you are not touching
the simulator beyond this point.
4.
Proceed to test instrument for leakage current as per established facility protocols
and procedures for safety testing of medical devices.