SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-9
Status Messages: Possible Causes:
Corrective Actions:
Sensor Invalid [0096]
1.
Poor or contaminated
connection between
sensor and PIC.
2.
An incorrect or
defective sensor has
been connected to the
PIC.
3.
Defective PIC.
4.
Defective BISx.
1.
Connect/clean connection
between sensor and PIC.
2.
Replace the sensor.
3.
Replace the PIC.
4.
Replace BISx Bulkhead
Connector or replace BISx.
Sensor used for over
24 hours
[ 109 ]
Sensor was attached to
system for more than 24
hours.
Replace sensor.
Unrecoverable
Monitor Error
[471]
Monitor was unplugged
during upgrade, or
battery ran out unexpectedly
and software did not have
time to detect it (check error
log for low battery entries)
1.
Unplug BISx and power
cord and turn monitor off.
2.
Reconnect AC power and
restart monitor.
3.
Press the reset button.
4.
If error persists, replace
monitor.
Data export error
[ 512 ]
(See Message 3000)
Date and time not
changed
[ 513 ]
A case was in progress when
the time/date was changed.
Change the date/time after the
case has finished.
Trend memory error
[ 516 ]
Trend memory is full or
data has been corrupted.
Disconnect the BISx and
restart monitor. Do not
reconnect the BISx until
directed.
Unrecoverable BISx
error
[ 517 ]
(See Message 1000)
No data available
[ 524 ]
1.
No data has been
written to the
BISx/monitor.
2.
Defective BISx.
3.
Defective monitor.
1.
Try again after a case has
been run.
2.
Replace BISx.
3.
Replace monitor.