SECTION 8 SERVICING THE BIS VIEW MONITORING SYSTEM
8-5
8.6
Replacing the Clamp Shoe
Parts Required:
Tools Required:
Shoe Clamp Replacement Kit P/N 186-0217
Philips #2 and #8 screwdrivers
To replace the Clamp Shoe, you will need Philips #2 and #8 screwdrivers. The power
supply must be removed to access the Clamp Shoe screws. Follow the instructions below:
1.
Unplug A/C line cord from the BIS VIEW monitor.
2.
Lay the monitor screen-side-down on a scratch-free work surface so that the
Battery/Power Supply cover is accessible.
3.
Remove 4 screws from the Battery/Power Supply cover and remove the cover. The
power supply is located inside the cover.
4.
Unplug (squeeze and pull to remove) the power supply connector.
5.
Remove screws from the power supply bracket and remove the power supply.
6.
Remove the three screws that attach the Clamp Shoe to the back panel.
7.
Replace the Clamp Shoe and attach it to the back panel with the three new screws
provided in the kit (hand-tighten only).
8.
Insert power supply into the cover, lining up the A/C power receptacle with the
cutout in the cover.
9.
Replace the power supply bracket and install screws (hand-tighten only).
10.
Reconnect the power supply.
11.
Replace the cover and 4 screws (hand-tighten only) and reconnect the A/C power
cord.
Содержание BIS VIEW
Страница 2: ......
Страница 4: ......
Страница 10: ...iii ...
Страница 78: ...SECTION 8 SERVICING THE BIS VIEW MONITORING SYSTEM 8 16 ...
Страница 99: ......