SECTION 3 PRINCIPLES OF OPERATION
_______________________________________________________________________
3-10
3.3.7
Data Transfer and Software Updates
Three ports on the rear of the BIS VIEW monitor facilitate data transfer. The USB (Type A)
port is used to export data to a removable drive. It is also used to update the monitor and
BISx software. The USB (Type B) port is for the manufacturer’s use only.
BIS values and other EEG data may also be acquired from the monitor using the RS-232
serial port. To connect to a personal computer, please contact Aspect Medical Systems
Technical Service for instructions. (See back cover for contact information.)
WARNINGS
THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING
WITH THE EQUIVALENT SAFETY REQUIREMENTS OF THIS
EQUIPMENT MAY LEAD TO A REDUCED LEVEL OF SAFETY
OF THE RESULTING SYSTEM. CONSIDERATION RELATING
TO THE CHOICE SHALL INCLUDE:
•
USE OF THE ACCESSORY IN THE PATIENT VICINITY
•
EVIDENCE THAT THE SAFETY CERTIFICATION OF THE
ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE
TO THE APPROPRIATE IEC 60601-1 AND/OR IEC 60601-2-26
HARMONIZED NATIONAL STANDARD.
WHEN CONNECTING EXTERNAL EQUIPMENT (e.g., DATA
CAPTURE COMPUTER), THE SYSTEM LEAKAGE CURRENT
MUST BE CHECKED AND MUST BE LESS THAN THE IEC
60601-1 LIMIT.
NOTE:
When software is updated, all previously recorded data and monitor configuration settings
will be lost. Therefore, configuration settings should be recorded before the software update
is performed.
Содержание BIS VIEW
Страница 2: ......
Страница 4: ......
Страница 10: ...iii ...
Страница 78: ...SECTION 8 SERVICING THE BIS VIEW MONITORING SYSTEM 8 16 ...
Страница 99: ......