TD 92816EN
17-July 2012 / Ver. B
42
Programming and Installation Manual
Care Phone
Mains return
Determines if Analog Care Phone / GSM Care Phone will send a return of power,
when power is restored. If this feature is enabled, notice of return of power is
sent from Analog Care Phone / GSM Care Phone from 5 to 35 minutes after the
power returns. The factory default is <Not activated>.
Passive Alarm
Determines the time in hours before the passive alarm is sent. Possible values are
2 -254 hours. In practice, this means that if the value is set to 24, the first alarm
is sent after about 24 hours. The time can also be set to 0 which means that the
alarm is triggered directly for, e.g., door alarms. Factory default is 24 hours.
9.11 Extended
Has a submenu, which controls the status of all functions in the Care Phone.
>Presence< >Ready< >Action< >How to send< >Emergency< >Assistance<
>Reminder Time<>No. of reminders< >Prefix D< >Prefix N, V<
Presence, Ready and Action
These are used to send information to the alarm receiver that the staff have
arrived at the care recipient, has carried out an action and is ready for departure
from the care recipient. These functions can be activated or deactivated by
selecting <Activated> or <Not activated> in the list that appears. A finish signal
is triggered automatically, a certain number of hours after a presence indicator, if
it has not previously been registered manually (the number of hours is
programmable, the default is 8 hours).
How to send
Time Indicates whether >Presence<, >Action< and >Ready< should be sent
individually, together or together with the test alarm.
Emergency (Staff Alarm)
Provides opportunity for staff to sound the alarm if they feel threatened. They
then use a special alarm transmitter (personal transmitter with radio code
22222222) recognised by all Analog Care Phone / GSM Care Phone -apparatus in
the area which have the function activated, thus triggering the alarm. The
appliances that have triggered the alarm, indicate where the alarm has come
from and the necessary assistance can be routed to the right location.
Assistance
If this function is activated the assistance alarm can be sent. This occurs if the
personnel are marked as present on the Care Phone and someone presses the
usual alarm transmitter or the red button on the Care Phone. An emergency
assistance can then go out to other staff. This function is used when the staff are
on the premises of the care recipient and need extra help.
Reminder time
Allows automated calling to the alarm receiver unless the staff have come to the
care recipient, and have marked their presence within a certain time. During this
time, Analog Care Phone / GSM Care Phone is blocked for further medical
alarms, medical alarms with low battery, emergency alarms, door alarms, bed
alarms, carpet alarms and user defined alarms. Analog Care Phone / GSM Care
Phone can be programmed to send a reminder alarm, 4 to 254 minutes after the
medical alarm. Value 0 means that the function is deactivated.