Publication No. 980938 Rev. K
Model T940 User Manual
Astronics Test Systems
Programmable Channel Calibration 6-17
Run Calibration
Equipment: Basic Setup
Procedure:
1. Press the Run button. Use the Stop button at any time to abort
execution.
2. Review the results in the Status window.
3. (Optional) Verify the results using the Verify button. Insure that all
channels pass verification.
4. (Optional) Check the individual gain and offset values for CVH and
CVL in the field controls. These values are not in engineering units.
5. (Optional) Save the calibration to a file for later restore, e.g., File |
Load DRA Calibration.
6. (Optional) Update the module to the new calibration factors just
obtained using the Update button. If this step is omitted, the
calibration factors will revert at the next power cycle.